IRGASON

®

Integrated CO

2

/H

2

O Open-Path Gas Analyzer and 3D Sonic Anemometer

o Analog output: use cable CABLE4CBL-L (pn 21972) or

CABLE2TP-L (pn 26986-L). Once again, the customer specifies

the length of this cable at time of order. The connector labeled

Analog Outputs on the EC100 panel indicates where each wire

should be connected (CO

2

voltage signal, H

2

O voltage signal,

and two ground connections).

6. Wire power and ground (i.e., power reference) cable CABLEPCBL-L (pn

21969-L) to the EC100. Feed the cable through one of the cable port

openings in the base of the EC100 enclosure and attach the ends into the

green EC100 power connector (pn 3768). Plug the connector into the

female power connector on the EC100 panel. Ensure that the power and

ground ends are going to the appropriate terminals labeled 12V and

ground, respectively.

7. Connect the power cable to a power source. The power and ground ends

may be wired to the 12V and G ports, respectively, of a Campbell

Scientific datalogger or to another 12 Vdc source.

8. Once power is applied to the EC100, three status LEDs on the EC100

panel will illuminate. The power LED is green and the sonic and gas

LEDs are orange until the unit has warmed up and is ready to make

measurements at which time the LEDs become green. If, after several

minutes, the LEDs turn red, a diagnostic flag has been detected. Check the

individual diagnostic bits to determine the specific fault. Diagnostics may

be monitored using the Status Window of ECMon, the user interface

software included with the IRGASON (see Section 7, Settings), or with a

datalogger (see Section 10, Datalogger Programming). The diagnostics

may reveal that the unit needs service such as cleaning the optical

windows on the gas analyzer, clear the sonic transducers of ice or debris,

etc. (see Section 9, Maintenance).

7. Settings

Operation of the IRGASON can be customized by changing settings. Factory

defaults work well for most applications, but the user may adjust the settings

with a PC using either the Campbell Scientific ECMon software (see Section

7.14, ECMon) or DevConfig (see Section 7.15, Device Configuration Utility),

or with a datalogger using the

EC100Configure() instruction (see Section

10.2, EC100Configure() Instruction).

15