After an extended period of time in the field, the IRGASON gas analyzer may

need to undergo this factory calibration again to ensure valid measurements.

When recalibration is deemed necessary, contact Campbell Scientific.

The IRGASON sonic anemometer also undergoes an initial factory calibration.

Usually this calibration remains valid unless a transducer fails or damage to the

instrument leads to a change in geometry. The sonic anemometer requires

calibration when

it develops a wind offset greater than the specification.

it sets diagnostic flags under dry conditions with little to no wind and

with no obstruction in the ultra sonic paths.

The wind offset is tested by creating a zero wind environment by encircling the

sensor with a large plastic bag (unused refuse bag), using caution not to block

the sonic paths. See the manual CSAT3 Three Dimensional Sonic Anemometer,

Section 11.2.2, Test for Wind Offset, available at

www.campbellsci.com

. When

recalibration is deemed necessary, contact Campbell Scientific.

10. Datalogger Programming

The datalogger is programmed in the CRBasic language, which features two

instructions for communication with the EC100 via SDM. The first instruction

is

EC100(), which reads measurement data from the EC100. The second is the

EC100Configure(), which receives and sends configuration settings.

10.1 EC100() Instruction

The

EC100() instruction is used to retrieve data from the IRGASON via SDM.

The instruction syntax is:

EC100 (Dest, SDMAddress, EC100Cmd)

Dest is the input variable name in which to store data from the IRGASON.

The length (i.e., number of data elements) of the input variable array will

depend on the selected value for EC100Cmd. A value of -99999 will be loaded

into Dest(1) if a signature error on SDM data occurs.

EC100Cmd

Dest Variable Length

0

8

1

12

2

13

SDMAddress defines the address of the IRGASON with which to

communicate. Valid SDM addresses are 0 through 14. Address 15 is reserved

for the

SDMTrigger() instruction.

EC100Cmd is a parameter that requests data from the analyzer. The results for

the command are returned in the array specified by the Dest parameter. A

numeric code is entered to request the data, as shown in TABLE 10-1.