IRGASON

®

Integrated CO

2

/H

2

O Open-Path Gas Analyzer and 3D Sonic Anemometer

any integer value between 0 and 14. The value 15 is reserved as an SDM-group

trigger.

7.4 Unprompted Output

If the EC100 is to output data as USB or RS-485 (see Section 8.2, USB or

RS-485 Output), this setting must be set to

USB Port or RS485, respectively.

The factory default is to disable the unprompted output and assume data are to

be logged via SDM (see Section 8.1, SDM Output).

Only one unprompted-output type (i.e., USB or RS-485) may be selected at a

given time. The rate at which the EC100 outputs these data is determined by

the

Unprompted Output Rate setting.

7.5 Unprompted Output Rate

This setting determines the output rate for unprompted output (USB or RS-485;

see Section 8.2, USB or RS-485 Output). If the unprompted output is disabled,

this parameter is not used. The factory-default output rate is 10 Hz, but it may

be set to 10, 25, or 50 Hz.

7.6 RS-485 Baud Rate

If the unprompted output mode is set to RS-485, this parameter determines the

baud rate. Otherwise, this setting is not used. The RS-485 baud rate defaults

to 115200 bps, although the user may enter another value.

7.7 Analog Output

The EC100 has two analog outputs for CO

2

and H

2

O densities (see Section 8.3,

Analog Outputs, for more information). These outputs may be enabled or

disabled with this setting. The default is for analog output to be disabled. Note

that the analog output does not include sonic data.

7.8 ECMon Update Rate

This setting determines the rate at which data are sent over the USB connection

to the PC while running ECMon. The default setting of 10 Hz should be

adequate in most situations.

7.9 Temperature Sensor

This setting configures the EC100 electronics to integrate with the IRGASON.

The IRGASON measures ambient air temperature using a thermistor

temperature sensor. With

Auto-Select selected, the EC100 will detect that the

IRGASON is connected and will report ambient air temperature measurements

from the thermistor.

To diagnose problems with the temperature measurement, a fixed temperature

value may be used, or the temperature sensor may be selected manually.

7.10 Fixed Temperature Value

If the

Temperature Sensor setting is None, the IRGASON will use the value

of this setting for the sample temperature. This mode is intended for

troubleshooting only. In normal operation,

Temperature Sensor is set to

Auto-Select, and this setting is not used.