beautypg.com

Ec100 outputs, 1 sdm output, 2 usb or rs-485 output – Campbell Scientific IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer User Manual

Page 32: Sdm output, Usb or rs-485 output

background image

IRGASON

®

Integrated CO

2

/H

2

O Open-Path Gas Analyzer and 3D Sonic Anemometer

8. EC100 Outputs

The EC100 is the IRGASON electronics module that measures the sensors and

controls the system. It outputs data in one of four types: SDM, USB, RS-485,

or analog. In general, Campbell Scientific recommends that SDM be used if a

Campbell Scientific datalogger is responsible for data collection. However,

RS-485 output is recommended if cable lengths exceed 100 meters. If a PC is

being used as the collection vehicle, USB and RS-485 are suitable outputs.

Analog output may also be used; however, only CO

2

density and H

2

O density

will be output. Information regarding each output type is provided in the

following sections.

8.1 SDM Output

SDM (

Synchronous Device for Measurement) is a proprietary Campbell

Scientific communication protocol that allows synchronized measurement and

rapid communication between a Campbell Scientific datalogger and multiple

devices, including the IRGASON. Although nearly all Campbell Scientific

dataloggers support SDM, only the CR1000, CR3000, and CR5000 dataloggers

support the IRGASON.

To use SDM data output, connect an SDM-communications cable from the

EC100 (see Section 6.3, Wiring and Connections) to a CR1000, CR3000, or

CR5000 datalogger. On CR1000 dataloggers, the SDM protocol uses ports C1,

C2, and C3. These are multipurpose ports that are SDM-activated when an

SDM instruction is used in the datalogger’s program. On CR3000 and CR5000

dataloggers, the SDM protocol uses SDM-dedicated ports SDM-C1, SDM-C2,

and SDM-C3.

Each SDM device on the SDM bus must have a unique address. The

IRGASON has a factory default SDM address of 1, but it may be changed to

any integer value between 0 and 14 (see Section 7.2, Bandwidth).

The sample rate for SDM output is determined by the inverse of the datalogger

scan interval, as set by the user in the datalogger program. Data are output

from the EC100 when a request is received from the logger, that is, a prompted

output mode. The number of data values sent from the EC100 to the

datalogger is also set by the user in the datalogger program. CRBasic, the

programming language used by Campbell Scientific dataloggers, uses the

EC100() instruction to get data from an IRGASON. This instruction is

explained in detail under Section 10.1, EC100() Instruction.

8.2 USB or RS-485 Output

In contrast to the SDM output mode, which is prompted by a datalogger, data

can also be output from the EC100 via USB or RS-485 in an unprompted

mode. In this case, the EC100 sends out data without initiation from the

receiving device at a rate determined by the EC100. Only one unprompted

output type, USB or RS-485, may be selected at a given time. RS-485 output

is recommended if cable lengths exceed 100 meters. If a Campbell Scientific

datalogger is not being used to collect the data from the IRGASON, either

unprompted mode is recommended.

22