Appendix G. CRBasic Program Library

'Offset (shift) to be applied to sensor output frequency

Dim

Off(2)

=

{ 0.0, 0.0}

'Steinhart-Hart coefficients [A,B,C] for converting thermistor ohms to

'temperature in Celsius. Specifying zeroes for A,B,C results in a reading in Ohms.

Dim

SteinA(2)

=

{ 0.0, 0.0}

Dim

SteinB(2)

=

{ 0.0, 0.0}

Dim

SteinC(2)

=

{ 0.0, 0.0}

'Rainflow configuration (not used in this program,

'but required as configuration arguments)

Dim

RFMB(2)

As Long

=

{ 20, 20}

Dim

RFAB(2)

As Long

=

{ 20, 20}

Dim

RFLL(2)

=

{ 400.0, 400.0}

Dim

RFHL(2)

=

{4000.0,4000.0}

Dim

RFHY(2)

=

{ 0.005, 0.005}

Dim

RFOF(2)

As Long

=

{ 100, 100}

'Configure the CDM-VW300 series device

'Use the variable arrays declared above

CDM_VW300Config

(0,CPI_ADDR,0,Enable(),Max_AMP(),F_Low(),F_High(), _

OutForm(),Mult(),Off(), SteinA(),SteinB(),SteinC(), _

RFMB(),RFAB(),RFLL(),RFHL(),RFHY(),RFOF())

DataTable

(static,true,-1)

'Static Frequency reading (1 Hz output)

Sample

(2,StaticFreq(),IEEE4)

'Thermistor reading : Ohms or DegC

Sample

(2,Therm(),IEEE4)

'Standard Deviation of dynamic readings

'taken during the most recent second

Sample

(2,DynStdDev(),IEEE4)

EndTable

DataTable

(dynamic,true,-1)

'Dynamic Frequency (20 Hz output)

Sample

(2,Freq(),IEEE4)

'Diagnostic code for the current dynamic reading

Sample

(2,Diag(),IEEE4)

EndTable

BeginProg

'20 Hz/50msec scan rate

Scan

(50,msec,500,0)

CDM_VW300Dynamic

(CPI_ADDR,Freq(),Diag())

'Get dynamic readings

CallTable

dynamic

If TimeIntoInterval

(0,1,Sec)

Then

' Process static data only once per second

CDM_VW300Static

(CPI_ADDR,StaticFreq(),Therm(),DynStdDev())

'Get static readings

CallTable

static

EndIf

NextScan

EndProg

G.1.2 20 Hz Measurement Example — One CDM-VW305, Eight

Channels

'===20Hz-1Device8Ch_4-25===

'CR3000 datalogger

'CDM-VW305 vibrating-wire analyzer

'Program to read 20-Hz dynamic data from one CDM-VW305 analyzer measuring eight channels

'IMPORTANT -- Ensure that the CPI address coded on the following line matches the address

'reported for the attached analyzer in the DevConfig or DVWTool software.

Const

CPI_ADDR

=

1

'<<<<<<<<<<<<<

Public

Freq(8)

'dynamic frequencies

Public

Diag(8)

As Long

'diagnostic code

Public

StaticFreq(8)

'Static (1Hz output) frequencies

Public

Therm(8)

'Thermistor readings

G-2