beautypg.com

Campbell Scientific CDM-VW300 Series Dynamic Vibrating-Wire Analyzer System User Manual

Page 19

background image

CDM-VW300 Series Dynamic Vibrating-Wire Analyzers

Sensor resonant frequency

range:

TABLE 5-2. CDM-VW300/305

Sensor Resonant Frequency Range (Hz)

Sample Rate

Minimum Sensor

Frequency

Maximum Sensor

Frequency

20 290 6000

50 290 6000

100 580 6000

200

3

1150 6000

333.3

3

2300 6000

Effective resolution (precision):

TABLE 5-3. CDM-VW300/305

Effective Frequency Measurement Resolution

4

Sample Rate (Hz)

Noise Level (Hz RMS)

5

1 0.005

20 0.008

50 0.015

100 0.035

200

3

0.11

333.3

3

0.45

Sustained input voltage without

damage:

–0.5 to 7.1 V

Temperature measurement:

Temperature measurement is
available for sensors so equipped.
Excitation and half-bridge
measurement circuits are integrated
into the CDM-VW300 and CDM-
VW305 analyzers.

Accuracy:

0.15% of reading

6

Resolution:

0.002 Ω RMS @ 5 kΩ thermistor
resistance

Bridge resistor:

4.99 kΩ 0.1%

Excitation voltage:

1.5 V

Measurement rate:

1 Hz

9