beautypg.com

2 measurement rates, 3 integrated thermistor measurement, Measurement rates – Campbell Scientific CDM-VW300 Series Dynamic Vibrating-Wire Analyzer System User Manual

Page 14: Integrated thermistor measurement, 2. eight-channel cdm-vw305 wiring panel, Figure 4-2 show

background image

CDM-VW300 Series Dynamic Vibrating-Wire Analyzers

FIGURE 4-2. Eight-channel CDM-VW305 wiring panel

4.2 Measurement Rates

CDM-VW300 analyzers use patented techniques to measure sensors at rates
from 1 to 333.3 Hz. Systems that exclusively require rates slower than 1 Hz
should use a Campbell Scientific AVW200-Series Vibrating-Wire Spectrum
Analyzer.

FIGURE 4-3. Measurement speeds of the AVW200 and CDM-VW300

analyzers

4.3 Integrated Thermistor Measurement

Many vibrating-wire sensors include an integrated thermistor to allow for
temperature compensation of the vibrating-wire measurement. The CDM-
VW300 supports thermistor measurements and the associated temperature
compensation required in some applications.

4