Appendix b. sc-cpi datalogger to cpi interface, B.1 introduction, B.2 quickstart – Campbell Scientific CDM-VW300 Series Dynamic Vibrating-Wire Analyzer System User Manual
Page 75: B.3 overview, Appendix b, See appendix b

Appendix B. SC-CPI Datalogger to CPI 
Interface 
B.1 Introduction
The SC-CPI is designed to interface the CR3000, CR1000, and CR800 
dataloggers to a network of CDMs (Campbell Distributed Modules) using the 
CPI bus communications protocol. Only one SC-CPI interface is required per 
datalogger. The datalogger and the SC-CPI should be installed in the same 
enclosure. 
B.2 Quickstart
The following figure shows SC-CPI to CR3000, CR1000, or CR800 series 
datalogger connection wiring. Cable length should not exceed five feet. 
SC-CPI module connects to terminals C1, C2, C3 (not the 
CR3000 SDM-C1, SDM-C2, SDM-C3). 
NOTE
B.3 Overview
The SC-CPI allows Campbell Scientific CR3000, CR1000, and CR800 
dataloggers to interface to CDM modules. The datalogger communicates with 
the SC-CPI module using control ports C1, C2 and C3. 
There are two LED indicator lights on the SC-CPI module. The indicator lights 
show status of the link and give a visual indication that it is operating correctly. 
The LED closest to the CPI port shows status of the CPI bus. The LED closest 
to the terminal block shows status of the serial connection to the datalogger. 
Connection to the datalogger consists of connecting 12V and G terminals of 
one device to the other. Power can come from the datalogger, but may also be 
provided by another source. A ground connection to the datalogger is required 
for a reference for the data signals. A ground-wire connection from the ground 
B-1
