2. dvwtool and crbasic settings – Campbell Scientific CDM-VW300 Series Dynamic Vibrating-Wire Analyzer System User Manual
Page 29

CDM-VW300 Series Dynamic Vibrating-Wire Analyzers
Procedure:
1. Install DVWTool before connecting the SC-CPI or CDM-VW300 to the
PC. Reference Section 7.1.1, Software and Driver Installation.
2. Before proceeding, follow the procedure outlined in Section 6.1,
Laboratory-Mode Installation for all sensors connected to the CDM-
VW300. Record settings determined in DVWTool for later use in the
CRBasic program. The following work sheet (TABLE 6-2) lists DVWTool
settings and corresponding CRBasic instruction settings, and provides a
place to record settings for one sensor. Copy and fill in this worksheet for
each sensor to be connected.
TABLE 6-2. DVWTool and CRBasic Settings
DVWTool Setting
Corresponding CRBasic
Instructions and
Parameters Setting
Device Name
No corresponding setting
Scan()
Scan Rate
1
Interval
1
No corresponding setting
Units = msec
No corresponding setting
Buffer = 500
No corresponding setting
Count = 0
CDM_VW300Config()
Device Type (auto-detected)
DeviceType
CPI Address
CPIAddress
No corresponding setting
SysOptions
Channel X Enable check box
ChanEnable
Resonant Amplitude (V)
ResonAmp
Minimum Frequency (Hz)
LowFreq
Maximum Frequency (Hz)
HighFreq
Output Format (Hz or Hz
2
)
ChanOptions
Multiplier
Mult
Offset
Offset
Steinhart-Hart Thermistor
Coeff A
SteinA
Steinhart-Hart Thermistor
Coeff B
SteinB
Steinhart-Hart Thermistor
Coeff C
SteinC
Rainflow | Number of Mean
Bins
RF_MeanBins
19