beautypg.com

Campbell Scientific CDM-VW300 Series Dynamic Vibrating-Wire Analyzer System User Manual

Page 30

background image

CDM-VW300 Series Dynamic Vibrating-Wire Analyzers

Rainflow | Number of Amp
Bins

RF_AmpBins

Rainflow | Low Limit

RF_LowLim

Rainflow | High Limit

RF_HighLim

Rainflow | Minimum Change

RF_Hyst

RF_Form

Rainflow | Rainflow Form |
reset list

A

Rainflow | Rainflow Form |
total list

B

Rainflow | Rainflow Form |
form list

C

1

Scan rate is automatically set based on the datalogger CRBasic Scan() instruction

Interval parameter. Set the Interval parameter such that the desired CDM-VW300
scan rate is achieved. Relationship between Scan Rate and Interval is:

CRBasic Scan() instruction Interval = (1 / DVWTool Scan Rate) * 1000

3. When the laboratory-mode installation has been performed on all sensors,

click Disconnect in the DVWTool window. Disconnect or turn off power
to the CDM-VW300.

4. Assemble the datalogger and power supply using FIGURE 6-7, Field data-

acquisition system, as a guide. Do not turn power on until the system is
completely assembled.

5. A reliable data-acquisition system requires a reliable power supply. CDM-

VW300 analyzers consume more power than do many Campbell Scientific
products, so the power supplies often used with other data-acquisition
systems may not be adequate. Consult Section 7.4, Using Power Supplies,
for sizing guidance. For large systems, consult with a Campbell Scientific
application engineer.

6. The following figure illustrates the connections between the datalogger,

SC-CPI interface, and CDM-VW300. With this figure as a reference,
work through the following procedure.

20