beautypg.com

Appendix e. calculating measurement error, E.1 example error calculation: geokon strain gage – Campbell Scientific CDM-VW300 Series Dynamic Vibrating-Wire Analyzer System User Manual

Page 85

background image

Appendix E. Calculating Measurement
Error

When using a CDM-VW300 analyzer, the basic output of a measurement is
frequency (Hz) or frequency squared (Hz

2

). Vibrating-wire sensors are usually

designed with a linear relationship between wire tension and the phenomenon
being measured. Further, wire tension usually has a linear relationship to the
square of the resonant frequency (f

2

). Manufacturers typically provide a

formula for converting frequency to an engineering unit (U

E

) in the form,

U

E

= Kf

2

,

Eq.

1

where K is a constant determined from sensor specifications.

So, for a very small change in frequency, ∆f, change in output, ∆U

E

, is given

by,

∆U

E

= 2Kf∆f

Eq.

2

where Eq. 2 is obtained by taking the first derivative of Eq.1 with respect to f,
and f is the present frequency on the wire.

Use Eq. 2 together with noise levels, N (Hz RMS), found in TABLE 5-3,
CDM-VW300/305 Effective Frequency Measurement Resolution, to calculate
the effective resolution (R

e

) of a measurement. For a specific sample rate in

TABLE 5-3, use the corresponding value of N in place of ∆f. Eq. 2 can then be
expressed as,

R

e

= 2KfN

Eq. 3

E.1 Example Error Calculation: Geokon Strain Gage

When using a Geokon strain gage, constant K is given by,

K = GB / 1000

Where G is the gage factor 4.062 or 0.391, dependent on gage type, and B is a
batch factor (a number usually near 1, such as 0.97), given on the sensor
calibration sheet.

Substituting into Eq. 1 using engineering units of microstrain (µε):

µε = GBf

2

/ 1000,

The effective resolution, R

e

, of the output, or ∆µε, is:

∆µε = 2GBfN / 1000

or

∆µε = GBfN / 500,

E-1