Appendix g. crbasic program library, G-13 – Campbell Scientific CDM-VW300 Series Dynamic Vibrating-Wire Analyzer System User Manual

Page 105

background image

Appendix G. CRBasic Program Library

Dim

Enable(2)

As Long

=

{ 1, 1}

Dim

Max_AMP(2)

=

{ 0.002, 0.002}

Dim

F_Low(2)

=

{ 300, 300}

Dim

F_High(2)

=

{ 6000, 6000}

Dim

OutForm(2)

As Long

=

{ 0, 0}

Dim

Mult(2)

=

{ 1.0, 1.0}

Dim

Off(2)

=

{ 0.0, 0.0}

Dim

SteinA(2)

=

{ 0.0, 0.0}

Dim

SteinB(2)

=

{ 0.0, 0.0}

Dim

SteinC(2)

=

{ 0.0, 0.0}

'Rainflow-Histogram configuration

Dim

RF_mean_bins(2)

As Long

=

{ MBINS, MBINS}

'Mean Bins

Dim

RF_amp_bins(2)

As Long

=

{ ABINS, ABINS}

'Amplitude Bins

Dim

RF_Lo_lim(2)

=

{ 400.0, 400.0}

'Low Limit

Dim

RF_Hi_lim(2)

=

{4000.0,4000.0}

'High Limit

Dim

RF_Hyst(2)

=

{ 0.005, 0.005}

'Hysteresis

'Output format consists of three digits ABC, that can be either 1 or 0

Dim

RF_OutForm(2)

As Long

=

{ 110, 110}

'RF Histogram Digit Code Histogram output result

'----------------------- ------------------------

'A = 0 Reset histogram after each output.

'A = 1 Do not reset histogram.
'B = 0 Divide bins by total count.

'B = 1 Output total in each bin.

'C = 0 Open form. Include outside range values in end bins.

'C = 1 Closed form. Exclude values outside range.

CDM_VW300Config

(0,CPI_ADDR,0,Enable(),Max_AMP(),F_Low(),F_High(), _

OutForm(),Mult(),Off(), SteinA(),SteinB(),SteinC(), _

RF_mean_bins(),RF_amp_bins(),RF_Lo_lim(), _

RF_Hi_lim(),RF_Hyst(),RF_OutForm())

DataTable

(static,true,-1)

Sample

(2,StaticFreq(),IEEE4)

Sample

(2,Therm(),IEEE4)

Sample

(2,DynStdDev(),IEEE4)

EndTable

DataTable

(dynamic,true,-1)

Sample

(2,Freq(),IEEE4)

Sample

(2,Diag(),IEEE4)

EndTable

'Store rainflow-histogram output

DataTable

(rainflhist,true,-1)

RainflowSample

(RF1(),IEEE4)

RainflowSample

(RF2(),IEEE4)

EndTable

BeginProg

'50 Hz/20msec scan rate

Scan

(20,msec,500,0)

CDM_VW300Dynamic

(CPI_ADDR,Freq(),Diag())

CallTable

dynamic

If TimeIntoInterval

(0,1,Sec) Then

CDM_VW300Static

(CPI_ADDR,StaticFreq(),Therm(),DynStdDev())

CallTable

static

EndIf

NextScan

SlowSequence

Scan

(5,sec,10,0)

'Rainflow-histogram call must be in a slow sequence scan

CDM_VW300Rainflow

(CPI_ADDR,RF1(),RF2())

CallTable

rainflhist

NextScan

EndProg

G-13