Appendix C. Public Variables
TABLE C-1. Public Variables
Usage
Variable Name
Units
Description
INFO
seq_ACTIVE
Boolean flag: True if the valve sequence is active
INFO/USER valve_number
valve number (1 to 11); determines which inlet is selected
INFO
sequence_index
index for valve switching sequence; cycles from 1 to
LEVELS_USED
INFO
smpl_counter
number of scans since last valve switch
INFO
end_omit
Boolean flag: True when equilibration time after valve
switch is satisfied
DIAG
panel_tmpr
°C
temperature of the datalogger wiring panel
DIAG
batt_volt
V
supply voltage; measured inside datalogger after diode
voltage drop
DIAG
batt_volt_LOW
Boolean flag: True if batt_volt is too low (controls power
shutdown)
DIAG
valve_tmpr
°C
temperature of the valve manifold
DIAG
valve_tmpr_OK
Boolean flag: True if the valve module is within operating
limits
DIAG
valve_heat_ON
Boolean flag: True if the valve heater is on
DIAG
valve_fan_ON
Boolean flag: True if the valve fan is on
USER
pump_ON
set = to True to enable the sample pump
DIAG
pump_press
kPa
pressure measured at the pump inlet
DIAG
pump_control
pump speed control variable; can be from 0 (off) to 1 (full
speed)
DIAG
pump_speed
Hz
pump speed; measured by a tachometer on the pump
DIAG
pump_tmpr
°C
temperature of the pump
DIAG
pump_tmpr_OK
Boolean flag: True if the pump temperature is within its
operating range
DIAG
pump_heat_ON
Boolean flag: True if the pump heater is on
DIAG
pump_fan_ON
Boolean flag: True if the pump fan is on
INFO
intake_heat_ON
Boolean flag: True if the intake heaters are on
TAIR
T_air(1)
°C
air temperature profile (optional, defined only if
N_AirTemps > 0)
TAIR
T_air(2)
°C
air temperature profile (optional, defined only if
N_AirTemps > 1)
TAIR
T_air(3)
°C
air temperature profile (optional, defined only if
N_AirTemps > 2)
TAIR
T_air(4)
°C
air temperature profile (optional, defined only if
N_AirTemps > 3)