beautypg.com

Campbell Scientific AP200 CO2/H2O Atmospheric Profile System User Manual

Page 74

background image

Appendix B. AP200 Diagnostics

Also check the measured pump temperature,

pump_tmpr and compare it to

the operating range (0°C to 55°C). The pump module has a heater that turns on

if

pump_tmpr falls below 2°C. If the pump temperature is too low, check the

operation of the heater which is controlled by public variable

pump_heat_ON.

This variable is saved in

RawData only if saving all diagnostics. Its

corresponding variable

pump_heat_Avg is saved in the averaged output tables

(

IntAvg, CalAvg, and SiteAvg).

The pump module has a fan that turns on if

pump_tmpr rises above 50°C. The

fan will stay on until the pump temperature falls below 45°C. If the pump

temperature is too high, check the operation of the fan which is controlled by

public variable

pump_fan_ON. This variable is saved in RawData only if

saving all diagnostics. Its corresponding variable,

pump_fan_Avg, is saved in

the averaged output tables (

IntAvg, CalAvg, and SiteAvg).

If the fan is on and the pump temperature is too high, check the ambient

temperature. The AP200 is rated for ambient temperatures from –30°C to

45°C.

Bit 4:

Pump pressure is not at the setpoint

If bit 4 of

diag_AP200 is set, this indicates the pump is not able to control the

pressure at the setpoint. To confirm the problem, compare the value of

pump_press to PUMP_P_SETPT. Diagnostic bit 4 indicates they differ by

more than 1.0 kPa.

Compare

pump_press (pressure measured at the pump inlet) to cell_press

(pressure measured by the IRGA). These two points are physically connected

by a tube with relatively low flow such that they should be at similar pressures.

The measured pressure should agree within the combined uncertainty of the

respective pressure sensors. If they disagree by more than 4 kPa, turn the

pump off (set

pump_ON = False) and allow the system to stabilize at ambient

pressure. Compare each pressure sensor to the pressure expected for the given

elevation. This test may help to diagnose a problem with the pump’s pressure

sensor. Also see the notes on bit 7.

If the pump pressure is too high, check

pump_control. If pump_control = 0,

this indicates the AP200 has turned the pump off. There are several reasons

the AP200 may shut the pump off:

Low supply voltage (see notes on bit 1)

Valve temperature out of range (see notes on bit 2)

Pump temperature out of range (see notes on bit 3)

Pump has been turned off by the user; if public variable

pump_ON =

False, set pump_ON = True to turn the pump on

If the pump pressure is too high and

pump_control is > 0, check the value of

pump_speed to verify the pump is running. If there is physical access to the

AP200, listen for the sound of the pump to confirm it is running.

If the pump pressure is too high and the pump is running, check the system for

leaks. If no leaks are found there may be a problem with the pump. Contact

Campbell Scientific for information on replacing the pump.

B-4