beautypg.com

6 verify performance, 1 quick status check using a keyboard display, Verify performance – Campbell Scientific AP200 CO2/H2O Atmospheric Profile System User Manual

Page 50: Quick status check using a keyboard display

background image

AP200 CO

2

/H

2

O Atmospheric Profile System

Sequence Timing, for timing details). When the sequence is restarted, the

AP200 will set

seq_ACTIVE = True and STARTsequence = False.

STOPsequence: Allows the user to stop the sequence. Set this variable to

True to stop the automatic valve switching sequence. The AP200 will

immediately set

seq_ACTIVE = False and leave the valve_number in its

current state.

valve_number: Controls which valve is selected. It may be 1 through 8

(profile inlets), 9 (zero), 10 (CO

2

Span), or 11 (H

2

O Span). If the sequence is

active (

seq_ACTIVE = True) the AP200 controls the state of valve_number

and the user may not change it. If the sequence is stopped (

seq_ACTIVE =

False), the user may change the value of valve_number to control the valves

manually.

5.6 Verify Performance

After the AP200 is installed, it is important to make sure it is operating

properly.

5.6.1 Quick Status Check Using a Keyboard Display

The optional, hard-mounted keyboard display or a hand-held CR1000KD

provides an easy user interface to the AP200. See Appendix A, Keyboard

Display Menu, for a listing of the AP200 keyboard display menus. See the

CR1000KD User Manual for operating instructions. Press a key to activate the

display and select the

Check Status page:

This menu page gives a quick look at the following public variables:

diag_AP200: this should be zero. If not, see Appendix B, AP200 Diagnostics.

CO2: this is the CO

2

concentration given in ppm. Verify that this is a

reasonable value.

H2O: this is the H

2

O concentration given in ppt. Verify that this is a reasonable

value.

seq_ACTIVE: this should be True. If not, see Section 5.5, Starting and

Stopping the Sequence.

38