H.2 common replacement parts – Campbell Scientific AP200 CO2/H2O Atmospheric Profile System User Manual
Page 106

Appendix H. Using Swagelok
®
Fittings
1. Do not remove the nuts and ferrules from the fitting. Simply insert
the tube into the assembled fitting until it bottoms out.
2. Rotate the nut until it is finger tight.
3. While holding the fitting body steady, tighten the nut until it feels
tight. Normally, this will be less than one full turn. Tightening a full
one and one-quarter turns will damage the threads on the fitting and
nut.
Reassembly, plastic or metal tubing:
Swagelok
®
tube fittings may be assembled and disassembled many times, but
the assembly process is slightly different than the first assembly.
1. Insert the tube with pre-swaged ferrules into the fitting until the front
ferrule seats against the fitting body.
2. Rotate the nut until it is finger tight.
3. While holding the fitting body steady, tighten the nut slightly with a
wrench.
H.2 Common Replacement Parts
Tubing
Campbell Scientific can provide several types and sizes of plastic tubing. A
tubing cutter, pn 7680, can be used to cut these tubes.
TABLE H-1. Available Plastic Tubing Sizes, Construction, and Usage Guidelines
CSI pn Tubing Type OD (in) ID (in)
Length (ft)
Construction
Notes
15702
Synflex 1300
1/4
0.17
500
Black HDPE
jacket, overlapped
aluminum tape,
ethylene
copolymer liner
Aluminum
layer limits
diffusion; best
for sample
tubes
15703
3/8
1/4
250
19164
1/2
3/8
250
26506
LLDPE
3/8
1/4
500
Black, linear, low-
density
polyethylene
More flexible
than HDPE
25539
1/2
3/8
500
19499
HDPE
5/8
1/2
100
Black, high-
density
polyethylene
Required for
larger
diameter
Tubing inserts
Inserts are recommended for use in plastic tubing. These inserts become
permanently attached to the tubing at the first assembly, so spare inserts may
be needed for replacing the ends of tubing.
H-2