beautypg.com

5 restart the sequence, 6 check the system, Maintenance and troubleshooting – Campbell Scientific AP200 CO2/H2O Atmospheric Profile System User Manual

Page 58: 1 plumbing leaks, Restart the sequence, Check the system, Plumbing leaks

background image

AP200 CO

2

/H

2

O Atmospheric Profile System

To avoid pulling unfiltered air into the AP200, do not shut down

or disconnect the dewpoint generator until the valve sequence

has been restarted as described in Section 6.2.5, Restart the

Sequence.

Press

to return to the Manual Zero/Span menu.

6.2.5 Restart the Sequence

At the conclusion of the manual zero/span process, restart the automatic valve

sequence. Go to the last line of the

Manual Zero/Span menu and set

STARTsequence = True. Verify valve_number changes to 1 and look at the

LEDs on the valve module to confirm that valve 1 is active. The AP200 will

stay on valve 1 until it is synchronized with the real-time clock. Then it will

set

seq_ACTIVE to True and begin the valve sequence. See Section 5.5,

Starting and Stopping the Sequence, for more details on starting the sequence.

Once the

valve_number changes to 1, the dewpoint generator may be shut

down and disconnected. Replace the plug on the AP200 H

2

O Span inlet.

Press

to return to the AP200 menu.

6.2.6 Check the System

When the manual zero/span process is complete and the valve sequence is

restarted, select

Check Status from the main AP200 menu and verify proper

operation of the system, as described in Section 5.6.1, Quick Status Check

Using a Keyboard Display.

7. Maintenance and Troubleshooting

Most of the basic diagnostic and troubleshooting issues for the AP200 are

covered in Appendix B, AP200 Diagnostics. This section provides additional

detail on some issues.

7.1 Plumbing Leaks

Most of the AP200 plumbing operates at reduced pressure. If there is a leak,

ambient air will be pulled into the air stream, mixing with the intended air

sample and changing the concentration of CO

2

and H

2

O. A large leak can be

easily identified by the higher-than-normal flow rate, but a small leak may be

difficult to detect. One technique that may be helpful to check for leaks is to

breathe on a suspect fitting while observing the measured CO

2

concentration.

Because exhaled breath contains much higher CO

2

than ambient air, breathing

on the location of a leak will cause a rise in measured CO

2

. If near the sample

inlet, take care that a user’s exhalations do not reach the inlet directly. There

may be a substantial time delay when checking for leaks at the intake

assembly. The mixing volume can cause the delay and smooth the response.

A long intake tube will also introduce a significant time delay.

NOTE

46