Address
:
7
>smode poll
(select the POLL mode)
Serial mode
: POLL
>reset
(resetting will activate the new ettings)
(no title will appear after reset in POLL mode)
send 7
(text invisible)
1013.25 hPa
The next picture (below) shows typical RS485/422 differential input
signals RS485/422 LO (upper signal) and RS485/422 HI (lower
signal) at baud rate 9600. The vertical scale is 2V/div for both signals.
The ground level for each signal is shown with a small arrow at the
right. At the left the signals are at rest.
The next picture (below) shows typical RS485/422 differential output
signals RS485/422 LO (upper signal) and RS485/422 HI (lower
signal) at baud rate 9600. The vertical scale is 2V/div for both signals.
The ground level for each signal is shown with a small arrow at the
right. At the left the signals are at rest.