Form defining the output format – Vaisala PTU200 User Manual

Page 17

background image

C

HAPTER

4_____________________________________________________________C

OMMISSIONING

V

AISALA

_________________________________________________________________________ 13

Examples:

>echo
Echo

:

ON

>
>echo off
Echo

:

OFF

FORM Defining the output format

FORM

The FORM command can be used to set or inspect the output format
of the transmitter. The maximum length of FORM is 80 characters.
The user can define the following fields into the output format:
amount of decimals

give number of decimals before a
quantity. Giving 4.2 before the
pressure quantity outputs a reading
with the following form: 1013.12

pressure quantities

P1, P2, P (average), PD
(difference)

pressure trend *)

TRE (three-hour trend)

pressure tendency

A (three-hour tendency)

temperature quantity of PT100
and HMP45D sensor heads *)

T

temperature quantity
of HMP45A-P *)

TH

relative humidity

RH

units

U, UU, UUU, UUUU, UUUUU

serial number of the transmitter

SN

error status

ERR (three characters)

stability indicator

OK (uses three characters)

checksums

CS2, CS4
CSX

number fields

n.m where: n = 0 - 9, m = 0 - 9

text fields

within “ “ characters

CR

\ r or #r

LF

\ n or #n

TAB

\ t or #t

nnn ASCII code

\ nnn or # nnn (cannot be 000)

*) The PTU200 transmitter cannot output + sign for pressure trend or
temperature reading; a space is output instead.