Form defining the output format – Vaisala PTU200 User Manual
Page 17

C
HAPTER
4_____________________________________________________________C
OMMISSIONING
V
AISALA
_________________________________________________________________________ 13
Examples:
>echo
Echo
:
ON
>
>echo off
Echo
:
OFF
FORM Defining the output format
FORM
The FORM command can be used to set or inspect the output format
of the transmitter. The maximum length of FORM is 80 characters.
The user can define the following fields into the output format:
amount of decimals
give number of decimals before a
quantity. Giving 4.2 before the
pressure quantity outputs a reading
with the following form: 1013.12
pressure quantities
P1, P2, P (average), PD
(difference)
pressure trend *)
TRE (three-hour trend)
pressure tendency
A (three-hour tendency)
temperature quantity of PT100
and HMP45D sensor heads *)
T
temperature quantity
of HMP45A-P *)
TH
relative humidity
RH
units
U, UU, UUU, UUUU, UUUUU
serial number of the transmitter
SN
error status
ERR (three characters)
stability indicator
OK (uses three characters)
checksums
CS2, CS4
CSX
number fields
n.m where: n = 0 - 9, m = 0 - 9
text fields
within “ “ characters
CR
\ r or #r
LF
\ n or #n
TAB
\ t or #t
nnn ASCII code
\ nnn or # nnn (cannot be 000)
*) The PTU200 transmitter cannot output + sign for pressure trend or
temperature reading; a space is output instead.