beautypg.com

General – Vaisala PTU200 User Manual

Page 53

background image

C

HAPTER

8_________________________________________________________ TECHNICAL DATA

V

AISALA

_________________________________________________________________________ 49

General

(* factory settings)
Temperature range

operating

-40 ... +60 ºC

with local display

0 ... +60 ºC

storage

-40 ... +60 ºC

with local display

-20 ... +60 ºC

Humidity range

non-condensing

Supply voltage

10 ... 30 VDC, reverse polarity
protected

Supply voltage sensitivity

negligible

Current consumption

less than 30 mA

with local display

less than 30 mA (without
backlight)
less than 50 mA (with backlight)

hardware shutdown mode

less than 0.1 mA

Serial I/O

full duplex RS 232C * or
bidirectional TTL level or
half duplex two-wire RS 485/422

code

ASCII

parity

even*, odd, none

data bits

7* or 8

stop bits

1* or 2

Baud rates

300, 600, 1200, 2400, 4800, 9600*

Electrical connections

excitation

supply, GND, CTRL (for
shutdown)

serial communication

TX, RX, GND, TXD, RXD, GND,
TXD inverted, RXD inverted,
GND, RS 485/422HI,
RS485/422LO

Units

pressure

hPa*, kPa, Pa, mbar, bar, inHg,
mmHg, torr, mmH

2

O, psia

humidity

%RH

temperature

°C*, °F

Resolution

0.1 hPa*, 0.1 %RH, 0.01 °C

Settling time at power-up

class A

class B

one pressure sensor

6 seconds*

5 seconds*

Acceleration sensitivity

negligible