Example:

T

ref

=

23.15°C

T

meas

= 23.28°C

Offset = -0.13°C

Gain

The gain correction can be calculated with the equation 6-2. It is
reasonable to use the low (T

1

) and the high (T

2

) end of needed

temperature range.

(

) (

)

2

1

2

2

1

1

1

meas

meas

meas

ref

meas

ref

T

T

T

T

T

T

Gain

+

=

(6-2)

where:
T

ref1

= measured with a reference thermometer at temperature T

1

T

ref2

= measured with a reference thermometer at temperature T

2

T

meas1

= measured with PTU200 at temperature T

1

T

meas2

= measured with PTU200 at temperature T

2

Example:
T

ref1

= -14.97°C and T

meas1

= -14.90°C

T

ref2

= +30.12°C and T

meas2

=+30.29°C

Gain = 0.99779

Offset = T

ref2

-Gain*T

meas2

= -0.10

T = 0.99779 * 30.29°C - 0.10 = 30.12°C

TXCI Entering corrections for HMP45D and PT100
temperature measurement

TXCI x

where:

x = 2 if there is one pressure transducer installed

3 if there is two pressure transducers installed

The TXCI command is used to enter new offset and offset/gain
corrections for temperature measurement of HMP45D and PT100.