Cald storing the date of calibration – Vaisala PTU200 User Manual
Page 41

C
HAPTER
6_______________________________________________
CALIBRATION AND
ADJUSTMENT
V
AISALA
_________________________________________________________________________ 37
correction ? -0.03
P1
6.
reading ? 947.42
correction ? -0.02
P1
7.
reading ? 1000.10
correction ? -0.04
P1
8.
reading ? 1099.58
correction ? -0.04
>mpc on
Multipoint adj:
ON
>
NOTE
Always remember to return the switch SW4 to write DISABLE
position (OFF).
CALD Storing the date of calibration
CALD yyyy-mm-dd
This command is used to store the date of calibration in the memory
of the transmitter.
Example:
>cald
Calibration date
????-??-??
>cald 1997-01-01
Calibration date
1997-01-01
>
Changing the status (ON/OFF) of the linear or multipoint corrections
or entering new linear or multipoint corrections will automatically
cancel the date of calibration.
NOTE
Always remember to return the switch SW4 to write DISABLE
position (OFF).