U

SER

'

S

G

UIDE

_______________________________________________________________________

28 ___________________________________________________________________ M210195

EN

-A

The POLL mode is used when several PTU200 series transmitters are
connected to one RS232C serial interface. It is also used when several
intelligent transmitters are connected to a half duplex RS 485/422
serial interface.

In the POLL mode, the transmitter must have a specific address so
that the host system is able to direct the outputting commands to the
particular transmitter. The addressable SEND command is used to ask
for a single message output from the transmitter.

The OPEN command can be used to open a single transmitter
temporarily to STOP mode. The CLOSE command will restore the
POLL mode.

SEND Outputting one single message

SEND a

where:

a = the address of the transmitter (0...99)

This command is used to output one message from the transmitter in
the POLL mode (see SMODE command).

Example:

>addr 7
Address

:

7

>smode poll
Serial mode

: POLL

>close

line closed

send 7

(text invisible)

1013.25 hPa