Humidity, Temperature, Hmp45a-p & hmp45d – Vaisala PTU200 User Manual
Page 52: Pt100 sensor head, Humidity temperature, Hmp45a-p & hmp45d pt100 sensor head

U
SER
'
S
G
UIDE
_______________________________________________________________________
48 ___________________________________________________________________ M210195
EN
-A
Humidity
Measurement range
0.8 ... 100 %RH
Accuracy at +20
°C (incl. non-linearity and hysteresis)
against factory references
±1 %RH (0...90 %RH)
±2 %RH (90...100 %RH)
field calibration against
references
±2 %RH (0...90 %RH)
±3 %RH (90...100 %RH)
Typical long-term stability
< 1 %RH / year
Temperature dependence
±0.05 %RH/ °C
Response time (90%)at +20
°C
15 s with membrane filter
Humidity sensor
HUMICAP
®
180
Temperature
HMP45A-P & HMP45D
Measurement range
-36 ... +60
°C (HMP45A-P)
-40 ... +60
°C (HMP45D)
Accuracy at 20°C
±0.2 °C
Accuracy over the whole measurement range:
0.1
0.2
0.3
-0.1
-0.2
-0.3
0
0.4
0.5
-0.4
-0.5
-20
°C
-10
0
10
20
30
40
50
60
-30
-40
Temperature sensor
HMP45A-P
Pt 1000 IEC 751 1/3 Class B
HMP45D
Pt 100 IEC 751 1/3 Class B
Pt100 sensor head
Measurement range
-40 ... +60
°C
Accuracy over
the whole measurement range
±0.2 °C
Temperature sensor
Pt 100 IEC 751 1/4 Class B