Chapter 8 technical data, Barometric pressure, Operating range – Vaisala PTU200 User Manual

Page 51: Accuracy class, Chapter 8, Technical data barometric pressure, Operating range accuracy class, Technical data

background image

C

HAPTER

8_________________________________________________________ TECHNICAL DATA

V

AISALA

_________________________________________________________________________ 47

CHAPTER 8

TECHNICAL DATA

Barometric pressure

Operating range

Pressure ranges

500 ... 1100 hPa, 50...1100 hPa

Accuracy class

Linearity *

± 0.05 hPa ± 0.10 hPa

Hysteresis *

± 0.03 hPa ± 0.03 hPa

Repeatability *

± 0.03 hPa ± 0.03 hPa

Calibration uncertainty **

± 0.07 hPa ± 0.15 hPa

Accuracy at +20 *C ***

± 0.10 hPa ± 0.20 hPa

Temperature dependence**** ± 0.1 hPa

± 0.1 hPa

Total accuracy including
one year drift

± 0.15 hPa ± 0.25 hPa

Long-term stability

± 0.1 hPa / year

Response time (100% response) class A

class B

one sensor

2 seconds

1 second

* Defined as the ±2 standard deviation limits of end-point non-
linearity, hysteresis error or repeatability error.
** Defined as ±2 standard deviation limits of inaccuracy of the
working standard including traceability to NIST.
*** Defined as the root sum of the squares (RSS) of end-point non-
linearity, hysteresis error, repeatability error and calibration
uncertainty at room temperature.
**** Defined as ±2 standard deviation limits of temperature
dependence over the operating temperature range.

± 0.20 hPa
± 0.08 hPa
± 0.08 hPa
± 0.20 hPa
± 0.30 hPa

± 0.30 hPa

± 0.45 hPa

± 0.20 hPa/year

Pressure range

500...1100 hPa

50...1100 hPa

class A

class B