beautypg.com

Switch sw4 in write enable position (on) – Vaisala PTU200 User Manual

Page 36

background image

U

SER

'

S

G

UIDE

_______________________________________________________________________

32 ___________________________________________________________________ M210195

EN

-A

(ON) to be able to enter new linear and multipoint corrections to the
transmitter.

ON

OFF

SW1

SW2

SW3

SW4

SW5

SW6

SW7

SW8

S1

F

IGURE

6-1

Switch SW4 in write ENABLE position (ON)

NOTE

Always remember to return the switch SW4 to write DISABLE
position (OFF).

CORR Listing linear and multipoint pressure corrections

CORR

The CORR command is used to indicate the status of linear and
multipoint pressure corrections. In case the status is ON, the
corrections and valid date of calibration are listed. The listing varies
according to the number of transducers in the transmitter. The
transmitter lists * in the second column if there exists only one
pressure transducer in the transmitter.

Example of the listing of a transmitter with one pressure transducer:

>corr
Linear adjustments

ON

Reading Correction

Reading Correction

500.010

0.120*******.**********.***

1100.320

0.150*******.**********.***

Multipoint adjustments

ON

Reading Correction

Reading Correction

499.660

-0.110*******.**********.***

599.110

-0.080*******.**********.***

698.580

-0.060*******.**********.***

800.950

-0.030*******.**********.***

900.400

0.010*******.**********.***

947.200

0.020*******.**********.***

999.840

0.050*******.**********.***

1099.090

0.070*******.**********.***