beautypg.com

R starting the continuous output, S stopping the output – Vaisala PTU200 User Manual

Page 30

background image

U

SER

'

S

G

UIDE

_______________________________________________________________________

26 ___________________________________________________________________ M210195

EN

-A

R Starting the continuous output

R

This command is used to start continuous outputting in the STOP and
RUN modes and after setting the output interval (see page and 18).

Example (in STOP mode):

>reset

PTU200 / 1.01
>r
1013.25 hPa 22.9 %RH 22.4 'C
1013.25 hPa 22.9 %RH 22.4 'C
1013.25 hPa 22.) %RH 22.4 'C
...

Example of restarting the outputting (in RUN mode):

>SMODE
Serial mode

: RUN

>RESET
1009.1 hPa 42.4 %RH 24.2 'C
1009.1 hPa 42.4 %RH 24.2 'C
...

S

(

text invisible)

>R
1009.2 hPa 42.5 %RH 24.2 'C
1009.1 hPa 42.5 %RH 24.2 'C
...

S Stopping the output

S

This command is used to stop the continuous outputting (activated
with power-up or reset in RUN mode or using the R command).

Example:

>r
1013.25 hPa 22.9 %RH 22.4 'C
1013.25 hPa 22.9 %RH 22.4 'C
1013.25 hPa 22.9 %RH 22.4 'C

s

(text invisible)

>