beautypg.com

Vaisala PTU200 User Manual

Page 33

background image

C

HAPTER

5________________________________________________________________

OPERATING

V

AISALA

_________________________________________________________________________ 29

OPEN Setting a transmitter from POLL mode to STOP
mode

CLOSE Setting a transmitter from STOP mode to POLL
mode

OPEN a CLOSE

where:

a = the address of the transmitter (0...99)

These commands are used to set a transmitter temporarily to STOP
mode and back to POLL mode again. The OPEN command is useful
when several transmitters are connected to one serial bus and
communication with a single transmitter is needed. The CLOSE
command is a global command and requires no address.

The PTU200 series transmitters must always be closed after the POLL
mode selection. The RESET command and powering-up will close the
PTU200 series transmitters automatically.

>addr 7
Address

:

7

>smode poll
Serial mode

: POLL

>close

line closed

Example of opening and closing a transmitter with address 7:

open 7

(text invisible)

PTB

7 line opened for operator commands

>

Any commands (except for the ? command) can now be used for the
transmitter with address 7 without affecting other transmitters with
different addresses connected to the same serial interface. However, if
the RESET command is given or the transmitter is powered-up, the
transmitter will automatically return to the POLL mode.
Example of closing a transmitter:

>close

line closed