Vers software version, Snum serial number, Errs error message output – Vaisala PTU200 User Manual
Page 49

C
HAPTER
7____________________________________________
SELF
-
TESTING AND PROBLEM HANDLING
V
AISALA
_________________________________________________________________________ 45
VERS Software version
VERS
The VERS command is used to output the software version of the
transmitter.
>vers
PTU200 / 1.01
>
SNUM Serial number
SNUM
The SNUM command is used to output the serial number of the
transmitter.
>snum
CPU
serial no : T2210005
TR1
serial no :
0
TR2
serial no :
0
TR3
serial no :
0
CTR1 serial no :
0
CTR2 serial no :
0
CTR3 serial no :
0
>
ERRS Error message output
ERRS
The ERRS command is used to print the error messages. The
command outputs an error code and an error description:
>errs
E00 Nothing special to report
>
Complete list of error codes:
E00 Nothing special to report
E10 CPU
EEPROM ackn. error
E20 CPU
EEPROM csum
error
E11 TR1
Transducer not present
E21 TR1
EEPROM csum error
E31 TR1
serial number error