Operating modes, Smode selecting the sending mode – Vaisala PTU200 User Manual
Page 15
C
HAPTER
4_____________________________________________________________C
OMMISSIONING
V
AISALA
_________________________________________________________________________ 11
Operating modes
Select the desired sending mode for the transmitter. This is done with
a command SMODE.
SMODE Selecting the sending mode
SMODE x
where:
x = STOP, RUN, SEND or POLL
The SMODE command is used to set or inspect the sending mode of
the transmitter. The PTU200 series transmitters have four sending
modes: STOP, RUN, SEND and POLL.
In STOP mode, after power-up the transmitter outputs its type and
software version and then waits for further commands.
In RUN mode, continuous outputting starts automatically from power-
up.
In SEND mode, a single message is automatically output after power-
up.
POLL mode allows the communication with multiple transmitters or
other digital instruments connected to one serial bus. The transmitter
does not echo in POLL mode.
Examples:
>smode
Serial mode
: STOP
>smode run
Serial mode
: RUN
>smode send
Serial mode
: SEND
>smode poll
Serial mode
: POLL
>smode stop
Serial mode
: STOP
>reset
NOTE
Remember to give the RESET command to initialize the new sending
mode.