beautypg.com

Lc linear pressure corrections – Vaisala PTU200 User Manual

Page 37

background image

C

HAPTER

6_______________________________________________

CALIBRATION AND

ADJUSTMENT

V

AISALA

_________________________________________________________________________ 33

Transducer 2

CORRECTIONS:

TXD0,TXD1,TXD2

:

-0.12000

1.00000

0.00000

V0XD0,V0XD1,V0XD2:

0.00000

1.00000

0.00000

V1XD0,V1XD1,V1XD2:

0.00000

1.00000

0.00000

Calibration date

1998-12-11

>

LC Linear pressure corrections

LC x

where:
x = ON or OFF
The LC command is used to activate or deactivate the linear offset or
offset/gain pressure corrections. Plain command LC outputs the linear
corrections in use.
The linear corrections are protected with switch SW4 (see

F

IGURE

6-1

)

Turn the switch SW4 to write ENABLE position (ON) to be able to
turn the linear corrections on or off.

NOTE

Changing the linear corrections on or off will automatically cancel the
date of calibration of the transmitter.

The listing varies according to the number of pressure transducers.
Example of linear adjustments performed on a transmitter with two
transducers:

>lc off
Linear adj.

:

OFF

>lc on
Linear adj.

:

ON

>lc
Reading Correction

Reading Correction

1013.250

-0.100

800.000

0.050

1013.250

-0.100

1000.000

-0.020

>

NOTE

Always remember to return the switch SW4 to write DISABLE
position (OFF).