Stopping continuous outputting, Outputting reading once, Assign an alias for the send command – Vaisala HMT330 User Manual
Page 98

User's Guide _______________________________________________________________________
96 ___________________________________________________________________ M210566EN-K
Stopping Continuous Outputting
Use the S command to end the RUN mode. After this command all other
commands can be used. You can also press the Esc button or reset the
transmitter to stop the outputting.
S
See command SMODE to change the default (power-up) operation
mode.
Outputting Reading Once
Use the SEND command to output the reading once in STOP mode.
SEND
Examples:
>send
RH= 98.4 %RH T= 31.1 'C
>send
RH= 98.4 %RH T= 31.1 'C Td= 36.0 'C Tdf= 36.0 'C a= 42.4
g/m3 x= 38.8 g/kg Tw= 30.8 'C ppm= 62414 pw= 59.53 hPa
pws= 60.52 hPa h= 130.7 kJ/kg
Assign an Alias for the SEND Command
Use the SCOM command to assign a new command that works like the
SEND command. The standard SEND command of the transmitter will
always function normally whatever the SCOM definition may be.
Command names are case-insensitive.
SCOM
Example (assigning the command MEASURE as an alias for the SEND
command):
>scom
Send command : ? measure
>measure
RH= 23.8 %RH T= 21.4 'C
>
You can remove the SCOM definition by pressing Esc when entering the
command name:
>scom
Send command : measure ?