164 __________________________________________________________________ M210566EN-K
Relative Humidity Adjustment after Sensor
Change
Using Display/Keypad
When using the optional display/keypad, follow the instructions in
section Using Display/Keypad on page 158 but select Adj. for new RH
sensor (instead of 1-point/ 2-point adjustment).
Using Serial Line
After sensor change, carry out the procedure as described in previous
sections. Just replace the CRH command with the FCRH command.
FCRH
Example:
>fcrh
RH
:
1.82 1. ref
?
0
Press any key when ready...
RH
:
74.22
2. ref
? 75
OK
>
The
OK
indicates that the calibration has succeeded.