Merit Medical PD Catheter Repair Kit IFU User Manual
Page 3

INSTRUCTIONS FOR USE
1. Create a sterile field for the procedure:
• Prepare the patient’s skin and existing catheter with a
cleaning or disinfecting solution as per hospital protocol.
• Drape the catheter exit-site area in an appropriate man-
ner.
2. If clamping the existing catheter or Repair Catheter
Tubing (C) with sterile hemostats or forceps, use only
smooth-jawed forceps or equivalent. Do not use forceps
with a serrated jaw; see cautions.
3. Prepare the existing Flex-Neck Peritoneal
Dialysis Catheter:
• Cut catheter tubing just distal to the existing connector
with sterile suture scissors, in a single, straight, perpen-
dicular cut. Do not pull the connector off the catheter.
• There must be at least 2.5 cm of catheter tubing remain-
ing after trimming.
• Verify that the cut is perpendicular to the tubing.
4. Select the appropriate Tube-to-Tube Connector, based
on the size of the existing catheter:
• Tube-to-Tube Connector (A) for all Flex-Neck Adult,
Adolescent and Pediatric Catheters
• Tube-to-Tube Connector (B) for all Flex-Neck Infant Cath-
eter
5. Flush and rinse the Repair Kit components with sterile
saline.
6. Wet the tapered tip of the appropriate Tube-to-Tube
Connector (A or B) with sterile saline or sterile water and
insert it into Repair Catheter Tubing (C).
• Do not use a twisting motion to force the Tubing onto
the Connector. Push the Connector into the Tubing with
a single forward motion.
• If Infant Tube-to-Tube Connector (B) is used, insert the
wider end of the Connector into Repair Catheter Tubing
(C). See Figure 2.
7. Advance the Repair Catheter Tubing (C) completely to
the raised ring barrier of Tube-to-Tube Connector (A) or (B).
See Figure 1 or 2, as appropriate. Pull carefully on Tube-to-
Tube Connector (A or B) and Repair Catheter Tubing (C) to
test the strength of the junction.
8. Wet the tapered tip of the Catheter Connector (D) with
sterile saline or sterile water, and insert it into the other
end of the Repair Catheter Tubing (C). Advance the tube
completely to the Connector’s raised shoulder ridge. The
catheter tubing must completely pass over the Tapered
Tip and to the Raised Shoulder Ridge, but not beyond that
onto the Finger Grip. See Figures 3 and 4. Pull carefully
on the Catheter Connector and Repair Tube to test the
strength of the junction.
• Do not use a twisting motion to force the Tubing onto
the Connector. Push the Connector into the Tubing with
a single forward motion.
9. Wet the remaining tapered tip of the Tube-to-Tube Con-
nector (A or B) in sterile saline or sterile water and carefully
insert it into the original, existing catheter. Advance the
catheter end completely to the raised ring barrier of the
Connector. See Figure 1 or 2, as appropriate. Pull carefully
on the catheter and connector to test the strength of the
junction.
• CAUTION: Do NOT dislodge the cuffs.
• Do not use a twisting motion to force the catheter onto
the Tube-to-Tube connector. Push the Tube-to-Tube
connector into the catheter with a single forward
motion.
10. Attach either Cap (E), or a dialysis transfer set, to the
threaded luer end of Catheter Connector (D).
CATHETER CLEANING AND CARE
All Flex-Neck Peritoneal Dialysis Catheters are made of
silicone. Exit-site cleaning agents that are compatible with
silicone catheters therefore may be acceptable for use on
Flex-Neck Peritoneal Dialysis Catheters and repaired cathe-
ters. Such cleaning agents include:
• Electrolytically-produced sodium hypochlorite solutions
(i.e., ExSept Plus®)
• Normal (sterile) saline
Cleaning agents that are non-irritating, non-toxic, anti-
bacterial, and in liquid form are generally recommended.
The following cleaning agents are not compatible with
silicone catheters, and are not recommended for use with
Flex-Neck Peritoneal Dialysis catheters:
• Acetone or acetone-based products
• Povidone-iodine or iodine-based products
Merit Medical Systems, does not provide specific recom-
mendations or protocols for exit-site care and cleaning,
whether by the healthcare professional or by the patient.
Appropriate exit-site and catheter care treatment protocols
should be individualized for each patient, and established
by the patient’s physician(s), nurse(s), dialysis center(s),
and/or other relevant dialysis healthcare professionals.
ExSept Plus is a registered trademark of Alcavis HDC, LLC or
one of its affiliates.
Copyright © Merit Medical Systems, Inc. All rights reserved.