When a process control relay channel is
configured to monitor Flow Total, it can be
used to dispense fluid in batches. A Process
Control Screen
for a channel configured in
this way is shown at right. The RELAY soft
button is used to visit each channel in turn
and go back to the Run Mode screen. The
START soft button is used to initiate a
batch. The left-most three soft button may
be used to change the Batch Amount or
clear the batch counter and the volume
totals.

8.2
Batch
Dispensing

EDIT

З

И

RELAY

START

Relay #3 Monitor: TOTAL

Status: ACTIVE

Batch Amount:

Current Batch #:

Total all Batches:

Batch Start:

Total this Batch:

Relay Timer:

5000

32

3296

156824526

AUTO

3

2 T 3.25 in.

[ ¢¢¢£]

1480

If the Batch Start type is configured as MANUAL, then operation is as follows:

!

Press the START soft button to begin the dispensing of a batch.

!

The START soft button is replaced by the STOP soft button.

!

The Total this Batch: value clears to zero.

!

The Current Batch #: value increases by one.

!

The relay is energized and the display changes accordingly.

!

As the flowmeter measures and totals up fluid flow, the Total this Batch: and Total

all Batches: values advance.

!

When the Total this Batch: value reaches the Batch Amount: value, the relay is de-

energized and the display changes accordingly.

!

The STOP soft button disappears and the START soft button returns.

A batch can be interrupted and resumed as follows:

!

Press the STOP soft button. The relay

is de-energized and the RESUM
(resume) soft button appears in place
of this soft button.

!

Although the display changes to show

that the relay is de-energized, it
alternates between black-on white and
white-on-black to indicate that a batch
has been interrupted.

!

The flowmeter is still measuring flow,

although the intent is clearly for the
flow to stop, and the operation of all
status displays continues.

!

Press the RESUM soft button. The

relay is energized again, the STOP soft button reappears and everything is as it
was before the interruption.

8.2.1
Manual Batch
Start

8.2.1
Manual Batch
Start Type