Proseries, U.s. gallons / min, Sonic-pro – Blue-White Sonic-Pro Hybrid Ultrasonic Flowmeters User Manual
Page 15
Page 15
Sonic-Pro
Industries, Ltd.
Industries, Ltd.
ProSeries
by Blue-White Ind.
TM
Separation
Distance
A
B
ACOUSTIC
COUPLING
MOUNTING CLAMPS
CENTERLINE
REAR VIEW
ARROW ON CENTERLINE
TRANSDUCER
BLUE GASKET - Permenant
installations. Recommended
for metal pipes.
ACOUSTIC COUPLING MATERIAL INSTALLATION
TRANSDUCER
BOTTOM VIEW
or
A
B
CENTERLINE
FRONT VIEW
ARROW ON CENTERLINE
CENTERLINE
A
B
An acoustic coupling material
must be placed between the
transducer and the pipe
surface at the point where the
sound waves enter the pipe.
Draw a centerline on the pipe
parallel to the pipe center.
Place the first transducer onto
the pipe. Locate the front
arrow exactly over the first
separation distance mark (A).
Place the transducer straight
and parallel to the centerline.
Secure the first transducer to
the pipe using the pipe clamps.
The arrows on the front and
rear of the transducers must
point toward the centerline.
Locate the second transducer
separation mark (B). Place the
second transducer so that the
front arrow is located exactly
over the mark (B). Place the
transducer straight and
parallel to the centerline. Be
sure that both transducers are
facing each other and parallel
to the pipe centerline.
Check that the separation
distance is correct. Tighten the
clamps equally.
7
Install the
transducers
CENTERLINE
ENTER
Ç
ESC
È
Setup Root Menu
Manage Configuration Sets
Global Configuration
8
Restart the
meter
Press the SETUP button. The
meter will enter the SETUP
ROOT MENU. Creation and
modification of the five
available Configuration Sets
and the Global Configuration
settings can be made from this
menu. If no changes are to be
made, press
ESC
. The meter
will re-start.
U.S. Gallons
U.S. Gallons / min
625.38
4345625.38
SETUP
SWAP
CLEAR
1 T 0.88 in.
[¢¢¢£]
1490
See section 6.9 for more information
DOW 111
SILICONE GREASE