Configuration Set #3

ÊData Logging

Metering Setup

Log Interval

К

З

180

К

Signal Output --> Analog Output allows you
to specify the mapping of flow rate to the
current value at the 4 to 20 mA analog
output signal. The mapping is specified by
two points on a straight line.

Signal Output --> Pulse Output allows you
to specify the mapping of flow rate to the
frequency of pulses at the digital output
signal. The mapping is specified by two
points on a straight line.

Data Logging --> Log Interval allows you to
specify a regular periodic interval in
seconds, at which flow rate and flow total
will be recorded in a log entry. The factory
default setting is 180 seconds. (see section
7.6 below for more on data logging).

Data Logging --> Log Rate Setpoint allows
you to specify upper and lower flow rates
which, when crossed, will result in log
entries being generated.

Data Logging --> Log Total Setpoint allows
you to specify five total flow values at which
log entries will be generated.

SOS Alarm --> Percent Error Window
allows you to define the percentage of fluid
sound speed change that is acceptable before
an error alarm is triggered. The default
value is 10%. See the index at the back of
this manual for fluid sound speed data.

ENTER

Ç

ESC

È

Configuration Set #3

Ê

Metering Setup

Analog Output

Pulse Output

Signal Output

Ê

ENTER

Ç

ESC

È

Configuration Set #3

ÊSOS Alarm

Metering Setup

Percent Error Window

К

10

К

К

ENTER

Ç

ESC

È

Configuration Set #3

ÊSOS Alarm

Metering Setup

Percent Error Window