Flow Total --> Volume Units allows

to

specify the units of measure in which flow
totals are displayed and written to log
entries. This is independent of the units of
measure for flow rates, so that the flow rate
units of measure mey be different than the
flow total units. As for Flow Rate, a custom
unit of measure can be specified.

Flow Total --> Digits After Decimal Point
allows

to specify how many digits

appear after the decimal point when flow
total is displayed on the Run Mode screen.

you

you

ENTER

Ç

ESC

È

Configuration Set #3

Ê

Metering Setup

Volume Units

Total Display Function

Digits After Decimal Point

Totalizer

Ê

Flow Total --> Total Display Function allows

to enable or disable the clearing of the

recorded total flow using the CLEAR soft button on the display in Run Mode and also
the clearing of batch counts and totals on Process Control Screens.

you

5.4.3
Metering
Setup
(continued)

higher accuracy and a more stable display but a slower response to changes in flow rate.
The Flow total value is not affected by flow rate averaging setting.

Display --> Display Update allows

to select the time interval between the

presentation of new flow information to the display, signal outputs, process control,
logging, etc. The choices offered are 0.25, 0.50, 1.00, 2.50 and 5.00 seconds. Both flow rate
and flow total are affected.

Display --> Low Flow Cutoff allows

to specify a flow rate below which the actual

measured value is replaced by zero. When factory configured, a flow rate equal to
approximately .25 feet per second is set.

Display --> High Flow Cutoff allows

to specify a high rate above which the actual

measured value is replaced by the cutoff value.

Display --> Scaling Offset allows

to specify a scaling factor, which would generally

be close to one, by which measured flow is multiplied. If required, this value can be used
to quickly adjust for errors in the flow rate reading after a field calibration has been
performed.

you

you

you

you

Display --> Display Language allows

to

select the natural language of display text.
The choices offered are English, Spanish,
German and French.

Display --> Flow Rate Averaging allows
to select the time interval over which flow
rate measurements are averaged before
presentation to the display, signal outputs,
process control, logging, etc. The choices
offered are 0.25, 0.50, 1.00, 2.50, 5.00 and
10.00 seconds. Longer values result in

you

you

ENTER

Ç

ESC

È

Configuration Set #3

Ê

Metering Setup

Display Language

Low Flow Cutoff

Display Update

Flow Rate Averaging

Display

Ê

High Flow Cutoff
Scaling Offset