Proseries, Sonic-pro, 3 metering setup (continued) – Blue-White Sonic-Pro Hybrid Ultrasonic Flowmeters User Manual
Page 35
Page 35
Sonic-Pro
Industries, Ltd.
Industries, Ltd.
ProSeries
by Blue-White Ind.
TM
Flow Total --> Volume Units allows
to
specify the units of measure in which flow
totals are displayed and written to log
entries. This is independent of the units of
measure for flow rates, so that the flow rate
units of measure mey be different than the
flow total units. As for Flow Rate, a custom
unit of measure can be specified.
Flow Total --> Digits After Decimal Point
allows
to specify how many digits
appear after the decimal point when flow
total is displayed on the Run Mode screen.
you
you
ENTER
Ç
ESC
È
Configuration Set #3
Ê
Metering Setup
Volume Units
Total Display Function
Digits After Decimal Point
Totalizer
Ê
Flow Total --> Total Display Function allows
to enable or disable the clearing of the
recorded total flow using the CLEAR soft button on the display in Run Mode and also
the clearing of batch counts and totals on Process Control Screens.
you
5.4.3
Metering
Setup
(continued)
higher accuracy and a more stable display but a slower response to changes in flow rate.
The Flow total value is not affected by flow rate averaging setting.
Display --> Display Update allows
to select the time interval between the
presentation of new flow information to the display, signal outputs, process control,
logging, etc. The choices offered are 0.25, 0.50, 1.00, 2.50 and 5.00 seconds. Both flow rate
and flow total are affected.
Display --> Low Flow Cutoff allows
to specify a flow rate below which the actual
measured value is replaced by zero. When factory configured, a flow rate equal to
approximately .25 feet per second is set.
Display --> High Flow Cutoff allows
to specify a high rate above which the actual
measured value is replaced by the cutoff value.
Display --> Scaling Offset allows
to specify a scaling factor, which would generally
be close to one, by which measured flow is multiplied. If required, this value can be used
to quickly adjust for errors in the flow rate reading after a field calibration has been
performed.
you
you
you
you
Display --> Display Language allows
to
select the natural language of display text.
The choices offered are English, Spanish,
German and French.
Display --> Flow Rate Averaging allows
to select the time interval over which flow
rate measurements are averaged before
presentation to the display, signal outputs,
process control, logging, etc. The choices
offered are 0.25, 0.50, 1.00, 2.50, 5.00 and
10.00 seconds. Longer values result in
you
you
ENTER
Ç
ESC
È
Configuration Set #3
Ê
Metering Setup
Display Language
Low Flow Cutoff
Display Update
Flow Rate Averaging
Display
Ê
High Flow Cutoff
Scaling Offset