0 setup (configuration) mode, Proseries, Sonic-pro – Blue-White Sonic-Pro Hybrid Ultrasonic Flowmeters User Manual
Page 27: 1 menu navigation, 1 setup root menu, 2 escaping from setup
Page 27
Sonic-Pro
Industries, Ltd.
Industries, Ltd.
ProSeries
by Blue-White Ind.
TM
5.1.1
Setup Root
Menu
The Setup (configuration) Mode screens are available on the Model S3 only. However, if
the communications option is included,
can access the
setup mode by using the PC Software Application.
any of the models S1, S2 or S3
5.1
Menu
Navigation
5.0 Setup (Configuration) Mode
The function of the soft buttons for menu
navigation is as follows:
Perform an appropriate action on
the highlighted item:
!
Select the item.
!
If an editable value, edit the value.
!
If the item has options, show the
options.
!
If the item is a selectable option, choose
it and move one place back up the menu
tree.
ENTER
Ç
ESC
È
Setup Root Menu
Manage Configuration Sets
Global Configuration
ENTER
Ç
ESC
È
ENTER
Ç
ESC
È
New Configuration Set
ÊFlow Rate
ÊVolume Units
Kilograms
ÊÇ Ounces Ç
U.S. Barrels Liquid
U.S. Barrels Oil
Pounds
È As Many Characters As Will Fit È
Metering Setup
The Setup Root Menu is shown at right.
From this screen, the user can Manage
Configuration Sets and edit certain
Global Configuration items.
The flowmeter stores up to five sets of
configuration item values, called
“Configuration Sets”, which are numbered
from 1 to 5.
5.1.2
Escaping From
Setup
Move the highlight up in the list of menu items or options.
Move the highlight down in the list of menu items or options.
Move one place back up the menu tree, abandoning any change at the current
position. If this soft button is pressed at the top level of a menu, leave the menu entirely,
abandoning all unsaved changes.
The function of the soft buttons and their labels will change when editing values, such as
numbers and passwords, and this is described in the following sections.
If the flowmeter enters Setup Mode and no input of any sort is received for a period of 5
minutes, then it will return to the Startup State, from which the user can re-enter the
Setup Mode, if desired.
If the ESC soft button is pressed repeatedly to go back up in a menu until the top level is
reached and is then pressed one time more, the flowmeter “escapes” from the Setup
Mode directly to the Startup State and from there to the Run Mode, as described in
section 4.4 above. The Configuration Set used is the same as was previously selected. If
no changes were made to this Set, for example if a different Configuration Set was
changed, the flowmeter will operate exactly as before. However, if the active Set has been
opened, edited and saved, then the changes will take effect without the re-calculation of
internal parameters and therefore the flowmeter may not operate correctly.