KROHNE BM 102 Handbook User Manual
Page 84

84
BM 102
 
To set the level theshold, we recommend the following filling height: 
• Minimum height: 300 mm or 12 ” above the probe tip or the upper weight, 
• Maximum height: distance to flange at least 500 mm or 20 “
 
Fct. Input distance 
This function is used to search for the level signal at a pre-configured distance. The current output 
is set to the appropriate value directly after the distance has been input. 
 
Note:
If the function is executed during normal operation, this may possibly have 
consequences on downstream systems. 
 
Fct. Search end of probe 
This function automatically defines the probe length. It is used, for example, after the probe has 
been shortened. The tank needs to be empty when this function is executed. If the probe tip is not 
found, possibly a level threshold has been set that is too high. Set the threshold to a lower value 
so that the reflection signal from the probe tip can be identified. 
 
Fct. Current output tests 
This function is used to check the accuracy of the analog output by means of a milliammeter. For 
this test a milliammeter must be connected into the loop. The following current values to be 
generated can be selected: 4/8/12/20/22 mA. 
 
Other functions:- 
F1 – Help 
On-line help file,
F3 – Exit
Exit to real-time measurement main window,
F10 – Screen copy
To select a directory and save the current screen as a bitmap file.
 
8.4.3 F3 
Exit:
To quit PCSTAR. 
 
8.4.4 F4 Serial (parameters): 
For configuring the gauge for networks. Refer to section 3.3.1 for a list of available parameters. 
 
8.4.5 F5 Record (Reading): 
Play back recording functions (offline). 
1.
The function will display a window: Record file – Open…. Browse for the DAT file 
required. Press OK to open file or Cancel to quit. 