Procedure (example where the level measured is too low compared to true level)
• read off displayed peak amplitude using the oscilloscope function – F7

• modify level threshold value


Step

Action

Data entered/ value
displayed

1

Execute PCSTAR 2. Press F2 to connect to gauge.

n/a

2

Press F11 to open F11-Dynamic configuration
window.

n/a

3

The top of the window gives the distance, peak
amplitude in volts and gain.

21000 mm
650 mV
Gain 3

4

The configured level threshold is listed under peak
amplitude in mV. Click on the “Modify” box to activate
the “Threshold” field.

n/a

5

Click on the ”Threshold” field and enter the new
value.

500 mV

6

If the approximate level is known, click on the
“Distance” field and then click on the “Search” button

-

7

If the value is still too low, try reducing threshold
again by small amounts. If this does not resolve the
problem, contact the KROHNE Service Centre. . A list
of service centres for each country is supplied on the
back page of this handbook.

n/a

Note that the threshold drops every 2 metres or 6.5 feet.



Solid application notes

Most dry solid applications except powder or flakes with high dielectric constants, such as coal
powder, are measured with a Gain of 3. If it is difficult to measure level at a certain point with the
gauge using a Gain from 0 to 2, then an internal tank structure (exposed girder, etc.) is most likely
to be within the electromagnetic pulse field area: the gauge will detect the largest signal and
assume this is the product level.