KROHNE BM 102 Handbook User Manual
Page 38
38
BM 102
Supplementary procedure for setting the 4 … 20 mA current to output volume readings
(using PCSTAR 2)
Step
Action
Data entered/value selected
1
Go to user function 1.3.1 ”Current 1 item” to select
measurement function.
Volume
2
Go to user function 1.3.2 ”Current 1 range” to set fault
status
4 … 20 mA
3
Go to user function 1.3.3 “Scale I1 min” to enter
volume value for the minimum output (4 mA)
0.50 m³ or 17.66 ft³
4
Go to user function 1.3.3 “Scale I1 max” to enter
volume value for the maximum output (20 mA)
16.80 m³ or 52.97 ft³
5
Either save the new configuration to disk by left-
clicking on F5-Save to disk and download it onto the
BM 102 using the F6-Send to BM102 key.
n/a
6
Press F3 to exit the configuration menu.
n/a
3.3.4 HART
®
Communicator: installation & operating instructions
Display and configuration can also be carried out with a HART
®
communicator. Operator control via
the separate HHC is described in detail in the operating instructions supplied with the HHC.
Basic instructions for taking measurements
Communicator layout
1
Two-pin jack for loop connectors
2 LCD
3
Function keys (F1 … F4)
4 Action
keys
On/off UP
Arrow
n/a
LEFT
Arrow
DOWN
Arrow
RIGHT
Arrow
Other functions:-
LEFT Arrow: Previous menu key
RIGHT Arrow: Select key