3.4.3 Level measurement when more than one phase or layer in the tank
Level can be measured with more than one phase in the tank. This requires setting a parameter in
the factory menu (Fct. 1.1.3: Application Type) to the following measurement mode:
• 2 liquids, 1 level

for measuring level with two or more phases

• 1 liquid, 1 level

for measuring one phase or liquid


Characteristics
• The level of the top product can be detected if it has a minimum layer of approximately

100 mm – when measuring a top product with a dielectric constant of 2.4.

• The mode “2 liquids, 1 level” permits level to be measured even when more than 2 liquids are

present in the tank.

• The first return signal is identified as being level and the second is ignored.

• This mode may be used with all probe types.


Example application 1: level measurement of oil (1 liquid in the tank)

With Factory Menu Fct. 1.1.3:
Application Type set to 1 liquid, 1
level, the BM 102 will search for the
return signal with the highest
amplitude (i.e. higher than the
threshold). It will measure the oil
level.

1


Level measurement signal

2 Threshold


Example application 2: level measurement when there are 2 liquids (oil/water) in the tank –
using a correctly configured gauge

With Factory Menu Fct. 1.1.3:
Application Type set to 2 liquids, 1
level, the BM 102 will search for the
first return signal higher than the
control threshold. It will correctly
measure the oil level (i.e. “level”).

1

Level signal

2

Threshold - finds first signal

3

Oil-water interface signal



For further information, please
contact your local KROHNE Service
Centre. Contact details are supplied
on the last page of this manual.