KROHNE BM 102 Handbook User Manual
Page 51

BM
102
51
 
Level measurement: Level pulse amplitude and threshold 
 
After connection to a power supply, the BM102 will: 
1. 
Measure reflection pulses in terms of voltage amplitude by cycling through a set of gains.
2.
Identify the highest amplitude as being the product level.
 
This screenshot of the PCSTAR 2 F7-Oscilloscope function while measuring a typical one product 
application identifies: 
1 Emitted 
pulse
2
Flange reflection (except coaxial probe)
3
Non-product reflection (e.g. parasite: agitator)
4
Product level reflection
5
Level threshold (with two-metre steps). Set in F11-Dynamic configuration menu
6
Offset (see below for definition)
7
Distance measured as a function of time
 
Offset is the distance for the signal to travel from the converter to the flange. 
(the gauge’s reference point). 
 
The level signal can be optimized by way of two factors: 
 
Amplification factor 
The amplitude of the signals is proportional to the dielectric constant of the product. At low 
amplitudes the signal should be amplified. The amplification factor is dependent on the dielectric 
constant and on the probe type. The device sets the gain automatically. 
 