KROHNE BM 102 Handbook User Manual
Page 45
BM
102
45
Function (Fct.)
Input Range
Description
2.2.0 TESTS
2.2.1 TEST OUTPUT
Select 4 mA, 20
mA, Other
This function allows the current
output to be tested.
The output can be set to one of the
listed values. With a reference
ammeter, the calibration of the
current output can be verified.
2.3.0 SERVICE
Restricted access factory
configuration menu. These
parameters may be accessed via
Fct. 3.2 “Specialist PSW
(password).
3.0 ACCESS
RIGHTS
Yes or no. Enter 9-
character code if
“Yes”.
3.1 MAINTENANCE
PSW
No
Disables the access lock on the
configuration menu.
The password must contain exactly 9
characters.
E, R or U are used only. The
password is displayed in a scrambled
format. It allows KROHNE to decode
the password in case it was forgotten.
3.2
SPECIALIST PSW
See KROHNE
Service centre or
BM 102 Service
Manual for code.
Specialist access to the Service
Parameters.
4.0 WATCH
STATUS
Displays the status of the device.
5.0 HART
®
VARIABLES
5.1 MANUFACTURER
KROHNE
Read only menu.
5.2 MODEL
BM 102
Read only menu.
5.3
FLD DEV REV
1.0
Field device revision.
Read only menu.
5.4 SOFTWARE
REV
1.0
Software revision.
Read only menu.
5.5 HARDWARE
REV
1.0
Hardware revision.
Read only menu.
5.6
DEVICE ID
Read only menu. The device ID is
also the serial number of the device.
5.7
MESSAGE
32 bytes of ASCII characters
5.8
DESCRIPTOR
16 bytes of ASCII characters.
Equivalent to PCSTAR2 Fct. 1.4.8
5.9
DATE
Month Day Year (xx / xx / xx).
5.10 NUM RESP PREAM
Number of preamble in the response
frame of the device.
5.11 TAG
Tag name of the BM 102.
5.12 POLL
ADDRESS
Address of the device.