KROHNE BM 102 Handbook User Manual
Page 79
BM
102
79
F3 – Exit
Quit to main title window.
F4 – Record Menu
The software will temporarily record the last two hours of operation without using the record menu.
This information will be displayed on the “Trend” and “Markers” windows. However, if the user
wishes to make a permanent record of measurement data then it is necessary to set and confirm a
measurement record period in this menu. It is also recommended to activate the “Oscilloscope”
option in order to have information on pulse reflection times.
This is set in the window below:
Functions available on this page:-
F1 – Help:
On-line help file
F2 – Start:
Start Record
F3 – Exit:
Exit to real-time measurement main window
Browse:
Find directory and create record file (*.dat)
Recording Options:
Start of Recording-
Either from now or a delayed start (date and
time)
End of Recording-
Either in a given number of hours or a given
date and time.
Signal/ Oscilloscope-
Activate these zones to record these types
of data.
Time slice-
Time spacing between each recorded
reading. Range: 1 to 30 minutes.
Comments-
Complementary information concerning the
record (50 characters maximum)