3.3.6 Local user display (instruments equipped with DIN connectors only)
The BM 102 local indicator is available as an option. This fits onto the BM 102’s standard DIN
connector. Data can be read from an LED display. The gauge configuration menu cannot be
accessed with this option. For non-Ex applications only. Please refer to the DA 06 Supplementary
instruction manual for further information.
Technical data: DA 06 Plug-in Display (non-Ex)
DA 06 Plug-in display (supplied separately) DA 06 Plug-in display mounted on BM 102 gauge
Measuring range:
-1999 to 9999 digits (min. and max. values user defined)
Display:
4-digit, 7.6 mm or 0.3” high, red LED display, display housing
can be turned through 330°
Accuracy:
0.1 % +/- 1 digit
Adjustable parameters:
Scaling, decimal point, attenuation, switching point, delay
Min. / max. value memory:
Displays the minimum and maximum values recorded during
operation.
Housing:
Plastic, PA 6.6 / polycarbonate 47mm X 47mm x 68 mm or 1.9” x
1.9” x 2.7” (h x w x d), housing can be turned by 300°
Protection:
IP 65 (EN 60529)
Temperature range (ambient)
Electronic system:
0°C … +60°C or 32°F … 140°F
Storage:
-30°C … +80°C or -22°F … 176°F
Electrical connection:
Adapter for plug as per DIN 43650
Input signal
4…20 mA / 2-wire
Max. voltage drop over unit
Approximately 6 V DC
Switching output1
open collector (PNP),
max. 125 mA,
on and off delay: 0 to 100 seconds,
temperature influence: 0.1% / 10 K
CE conformity
EN 50081-1 and EN 50082-2