Installation and operating instructions BM 70 A/P

Page: 4

11/00

Software history

Signal Converter Firmware

For BM 70 A:

Introduction
Mth./Yr

Firmware
Version

Device
Instructions

Remarks

06/97

3.00PREnn

06/97

Test versions

01/98

3.00

11/97-11/00

First serial version. Errors when empty spectrum recorded
with WAVE-STICK; automatic reset when signal is extremely
weak.

05/98

3.01

11/97-11/00

Elimination of errors in V. 3.00; improved application of empty
spectrum at tank bottom.

07/99

3.02

05/98-11/00

Implementation of the ModBus protocol; adaptation to
PROFIBUS interface; improved application of empty
spectrum at tank bottom; improvement of PC-CAT protocol
switching in multidrop operation; definition of 3.6 mA as lower
error current for the IS current output; improved tank bottom
tracing; improved spectrum evaluation at tank heights > 15
m.

11/99

3.03

05/98-11/00

Correction of a bug in the mode "Hold on Error" (output
current went to 1 mA or 3.6 mA; only V. 3.02); Adaption of the
HART protocol for operation with the program AMS, and
some smaller error corrections; Improved "tank bottom
recognition" (measured value now always goes to the
programmed tank bottom position); If no measured value is
found after start-up, the instrument goes directly into "tank
bottom recognition" and the local display to level = 0, if Fct.
3.2.5 Error Messages = NO is set.

For BM 70 P:

Introduction
Mth./Yr

Firmware
Version

Device
Instructions

Remarks

05/98

4.00PREnn
4.01PREnn
4.02PREnn

05/98

Test versions

07/99

4.02

05/98-11/00

First serial version.

11/99

4.03

05/98-11/00

Correction of a bug in the mode "Hold on Error" (output
current went to 1 mA or 3.6 mA; only V. 4.02); Adaption of the
HART protocol for operation with the program AMS, and
some smaller error corrections; Improved "tank bottom
recognition" (measured value now always goes to the
programmed tank bottom position); If no measured value is
found after start-up, the instrument goes directly into "tank
bottom recognition" and the local display to level = 0, if Fct.
3.2.5 Error Messages = NO is set.