Technical data – KROHNE BM 26 BASIC_ADVANCED EN User Manual
Page 55

TECHNICAL DATA
8
55
BM 26 BASIC/ADVANCED
www.krohne.com
10/2012 - 4000347004 - HB BM26 Basic/Adv R04 en
Display and user interface
Display and user interface
Display and user interface
Display and user interface
Display
None
Measuring accuracy
Accuracy
±
10 mm / 0.4¨ (when density is constant)
Operating conditions
Temperature
Temperature
Temperature
Temperature
Operating temperature,
transmitter module
-40…+85°C / -40…+185°F, if there is insulation around the
measuring chamber, specify the temperature in the order. Do not put
insulation around the transmitter housing.
Pressure
Pressure
Pressure
Pressure
Operating pressure
Atmospheric pressure
Other conditions
Other conditions
Other conditions
Other conditions
Protection category
Transmitter housing without LCD indicator: IP 54
Transmitter housing with LCD indicator: IP 66
Installation conditions
Notes
The analog transmitter is calibrated at the factory and attached to the
measuring chamber before delivery
Dimensions and weight
Refer to the "Technical data: Dimensions and weights" section
Materials
Housing
Polyester-coated aluminium
Reed-chain tube
Stainless steel
Clamp
Stainless steel
Electrical connections
Power supply
Power supply
Power supply
Power supply
Voltage
Non-Ex:
Non-Ex:
Non-Ex:
Non-Ex:
9...32 VDC
Ex ia:
Ex ia:
Ex ia:
Ex ia:
Refer to supplementary instructions or approval certificates
Cable entry
M20 × 1.5
Intrinsically-safe circuit data for
Ex ia-approved devices
Refer to supplementary instructions or approval certificates
PROFIBUS PA
PROFIBUS PA
PROFIBUS PA
PROFIBUS PA
Description
-
PROFIBUS PA protocol Profile
A&B, ver.3.0 (EN 50170 vol.2)
FOUNDATION Fieldbus
FOUNDATION Fieldbus
FOUNDATION Fieldbus
FOUNDATION Fieldbus
Description
FOUNDATION Fieldbus protocol
-
Module output
FOUNDATION™ Fieldbus
PROFIBUS PA
Order code
xF45xDxxxxx (PR)
xF45xXxxxxx (PR)