Analog transmitter

Optional LCD indicator status:

OL

Current output is more than 22 mA

(more than 20 mA, if this option is

selected in the indicator). Overfill.

The level of the liquid is above the

maximum limit (if the indicator is

configured for level indication).

Lower the level of liquid below the

maximum limit.

Current output is more than 22 mA

(more than 20 mA, if this option is

selected in the indicator). The

analog transmitter has been

moved. A value more than 20 mA

no longer represents the

maximum limit.

Contact your local sales office for

advice.

Current output is more than 22 mA

(more than 20 mA, if this option is

selected in the indicator). The

analog transmitter is defective. A

value more than 20 mA no longer

represents the maximum limit.

Contact your local sales office for

advice.

Optional LCD indicator status:

-OL

Current output is less than 3.6 mA

(less than 4 mA, if this option is

selected in the indicator). The level

of the liquid is below the minimum

limit (if the indicator is configured

for level indication).

Increase the level of liquid until it is

above the minimum limit.

Current output is less than 3.6 mA

(less than 4 mA, if this option is

selected in the indicator). The

analog transmitter has been

moved. A value less than 4 mA no

longer represents the minimum

limit.

Contact your local sales office for

advice.

Current output is less than 3.6 mA

(less than 4 mA, if this option is

selected in the indicator). The

analog transmitter is defective. A

value less than 4 mA no longer

represents the minimum limit.

Contact your local sales office for

advice.

The level of the liquid is at or

above the maximum level, but

the analog transmitter output

doe not indicate that level has

reached its maximum limit

yet.

The density of the liquid is lower

than the value given in the

customer order data. The float is

too low in the liquid.

Make sure that the density of the

liquid is stable. Contact your local

sales office for advice.

The analog transmitter output

indicates that the liquid is at

or below the minimum level,

but the level of the liquid is

still above the minimum limit.

The density of the liquid is lower

than the value given in the

customer order data. The float is

too low in the liquid.

Make sure that the density of the

liquid is stable. Contact your local

sales office for advice.

The level of the liquid is below

the maximum level, but the

analog transmitter output

indicates that level has

already reached its maximum

limit.

The density of the liquid is higher

than the value given in the

customer order data. The float is

too high in the liquid.

Make sure that the density of the

liquid is stable. Contact your local

sales office for advice.

The analog transmitter output

is not yet at the minimum

level, but the level of the

liquid is at or below the

minimum limit.

The density of the liquid is higher

than the value given in the

customer order data. The float is

too high in the liquid.

Make sure that the density of the

liquid is stable. Contact your local

sales office for advice.

Error

Description

Corrective action